top of page
Search
Janet Ferrao

Eppendorf 5810R 5804R compressor and fans test

Service Mode #6

Push the lid down and then press START

Push “temp “ button to turn the compressor on

CAUTION : THE COMPRESSOR WILL OVERHEAT IF YOU RUN IT WITHOUT RUNNING THE COMPRESSOR FAN.


To turn on the compressor fan push the “time” button

The Up and Down arrow keys are used to control the speed of the fan


Push the Program key to test the Blower fan

Push thr Speed key to test the electronic fans on the Frequency Converter PCB


19 views0 comments

Recent Posts

See All

Comments


bottom of page